2013 Aug 1 LCD Icons Representations Battery status indication icons These icons indicate the battery level. There are 3 levels representing empty, half and full. In dual-axis mode, the LCD displays the direction of tilt graphically. It will show as E bubble to display the direction of tilt Degree mode. Flash when unit is in HOLD mode mm/M, the height of one end for 1m long plate. Gradient % mode. Flash when unit is in HOLD mode Sound notification on. Blank as off Show this logo as relative value is showing. When the Zero button is pressed, the unit reset current angle to zero. Direction of tilt icons, show the tilt angle direction Dual-Axis Mode. Both the X and Y axis angle will be showed. Dual axis mode measures inclination up to +-40 degree for each axis before it automatically switch to single axis mode. Single Axis Mode. Measure slope up to +-90 degree. The triangle icon indicates the direction of tilt with respect to the bottom right corner of the unit. 4

*DXL360S/C Only

2013 Aug 1 Battery Charging

It has a built in Lithium Ion rechargeable battery. A standard charger is supplied that the input voltage is 110V to 240V AC, 50/60Hz, and the output is 5V DC, 500mA. The charger operating procedure is list below: 1) Plug the Charger into AC socket, the RED indicator on the charger should turned

ON,

2) Plug the USB charging cable to the Charger, 3) Insert the other end of the USB cable to the unit, 4) The battery icon on the LCD blinks to indicate charging in process. Upon charging

complete, the icon stops blinking.

5) The charging time is approx. 3 hours.

The unit could also be charged by connecting the USB cable to the unit and a computer’s USB port. This has the same effect when charging the unit with the provided adaptor. Note: When the unit is turned OFF, and plug in the USB charge cable, the LCD will no show anything, it is NORMAL. Once the unit is turned ON, the battery icon should flash indicating the unit is in charging mode.

Relative/ Absolute Measurement Relative Measurement LCD Icon:

Absolute Measurement LCD Icon: Blank

5

2013 Aug 1 Relative and Absolute Mode Switching:

1 Press and release the “ZERO” key to set the relative measurement zero

point 2 Press and release the “ZERO” key to cancel the relative zero point and

back to absolute measurement mode.

Hold Function Holding Mode: LCD Icon blinking:

Holding function:

1. Press and release the “HOLD” key to activate holding function, digit will

freeze. 2. Press and release the “HOLD” key to cancel the holding function.

Auto Power Off

For no movement in 30 minutes, the unit will power off. Or we can set to never sleep mode at below instruction. Power auto off setting:

1. Press and hold “SET” / “ANGLE” key and enter MENU mode

2. Select “POWER” by “ZERO” and “HOLD” key, press “SET” key to enter Power mode

3. Scroll “NEVER” or “30MIN” by “ON/OFF” and “HOLD” key

4. Press “SET” key to confirm NEVER or 30MIN (30 minutes) sleep

Restore Factory Setting

When you find that the unit is abnormal, you can restore the unit to factory setting.

6

2013 Aug 1 All calibration setting will be restored to factory setting.

*Not suggest restore to factory setting in normal status, for accuracy drift, please follow Calibration. After factory set, please redo calibration to ensure the accuracy. Restore to factory setting:

1. Press and hold “SET” / “ANGLE” key and enter MENU mode

2. Select “FACTORY SET” by “ZERO” and “HOLD” key, press “SET” key to enter FACTORY SET mode

3. Scroll “YES” or “NO” by “ZERO” and “HOLD” key 4. Press “SET” key to confirm

Angle Alarming Alarming Mode: LCD Icon:

Alarming Angle setting:

1. Press and hold “SET” / “ANGLE” key and enter MENU mode

2. Select “BUZZER” by “ZERO” and “HOLD” key,

press “SET” or “ANGLE” key to enter BUZZER setting

Press and hold “ZERO” and “HOLD” key for fast scrolling the digit. 3. Press “SET” or “ANGLE” to enter setting value

BUZZER >SINGLE : 0 0 . 0 0 DUAL. X: 0 0 . 0 0 DUAL. Y: 0 0 . 0 0 RANGE : 0 0 . 5 0 ACTIVE : IN OUT

7